Skip Navigation

This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow References
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Request Permissions
Google Scholar
Right arrow Articles by Babin-Ebell, J.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Babin-Ebell, J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?


Influence of valve size on morbidity and mortality for patients with mechanical valve replacements (GELIA 8)

J. Babin-Ebell*,1, C. Schuster1, O. Elert1, R. Bergemann2 and D. Horstkotte3 GELIA Study Group

a Klinik und Poliklinik für Herz- and Thoraxchirurgie, Universität Würzburg, Würzburg, Germany
b Institute for Medical Outcome Research, Lörrach, Germany
c Herzzentrum Nordrhein-Westfalen, Kardiologische Klinik, Ruhr Universität Bochum, Bad Oeynhausen, Germany

* Correspondence: PD Dr. med. Jörg Babin-Ebell, Klinik and Poliklinik für Herz- und Thoraxchirurgie der Universität Würzburg, Josef-Schneider-Str. 6, 97080 Würzburg, Germany.

Abstract

Aims In the past most of the investigations regarding the influence of valve size on outcome after heart valve replacements were done retrospectively. The present study is the first to investigate prospectively the influence of the implanted valve size on morbidity and mortality, using a close follow-up technique.

Method and results The findings in the present study were obtained prospectively in a multicentre study (the German Experience in Low Intensity Anticoagulation [GELIA] study) conducted from 1993 to 1999. In the subgroup analysis we sought to determine whether a smaller valve size influenced morbidity or mortality after heart valve replacement. Data from 2244 patients (1765 aortic and 479 mitral valve replacements) were analyzed for morbidity. For mortality, a total of 2557 patients were eligible. Patient data were compared for all valve diameters, and for valve area index (VAI) below and above the median (median VAI for aortic valves was 1·6 cmz. m-2 and for mitral valves it was 2·44 cm2 . m–2). Only three patients fell into the categories of small VAI (aortic valves VAI <0·9 cm2. m–2; mitral valves VAI <1·3 cm2. m–2).

Conclusion The present study was unable to demonstrate an influence of the size of the implanted heart valve on mortality and morbidity, either in the mitral or in the aortic position. The St. Jude Medical valve appears to provide a sufficient valve area, even for small sizes.

Key Words: Morbidity • mortality • valve replacement • valve size


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.